TEC 3000 Cryogenic Freezer Controller
The TEC 3000 employs a variety of advanced features that enable the controller to monitor and control the environment inside a freezer with a high level of precision.
Liquid Nitrogen Level Measurement
- The TEC 3000 uses a differential pressure system to determine
the LN2 level to an accuracy of± 0.5 in. (15 mm) and a resolution
ofO.l in. (5.0 mm). Unlike alternative level sensing systems,
differential pressure allows the exact level to be measured and
displayed. Using the simple single point calibration in a range
of 3.0 in. to 48.0 in. (75 mm to 1220 mm), the patented, selfmaintaining,
closed-loop system displays in inches, millimeters,
or a percentage full.
Automatic Liquid Nitrogen Level Control
- The fully automated LN2 level control system is based on userdefined
parameters that can be electronically adjusted over the
entire level range. The parameters include Low Level Alarm, Low
Level Fill Point, High Level Fill Point and High Level Alarm. The
redundant Dual Solenoid Valves for overfill protection run on 24
VDC, 1.0 amp (max).
Liquid Usage
- This exclusive feature provides an estimation of liquid usage to
track LN2 consumption and can provide an early failure warning
to allow sufficient time to implement corrective action and save
irreplaceable samples.
User-Defined Alarms
- A total of 17 audio/visual alarms are used to alert the user to any
potential or developing problems. The alarms include: High
Temperatures, Low Temperatures, High Level, Low Level, Liquid
Usage, Maximum Fill Time, Gas Bypass, Temperature Calibrations,
Low Battery, Power Failure, Lid Open and Communication
Loss.
Remote Alarm Monitoring
- The TEC 3000 Alarm monitoring includes a Global Remote
Alarm Relay, as well as four discrete contacts for High Level,
Low Level, Temp A High and Low Battery.
Temperature Measurement
- Two independent temperature measurement channels are employed
to accurately measure the temperature across the entire
storage space. The two platinum RTD sensors have an accuracy
of± l .0°C and a resolution of 0.1 °C. The temperature can be displayed
in °C, °F or K. The single or two point calibration also has
altitude compensation for the highest accuracy.
Hot Gas Bypass
- This unique feature is able to vent warm nitrogen gas from the
supply line before initiating a fill. This prevents warm gas from
entering the freezer space, which helps maintain a stable temperature
gradient and increases the efficiency by reducing excess
LN2 evaporation.
Event Log / Data Storage
- The TEC 3000 is able to store vital, unalterable, time-stamped
data in nonvolatile memory. This is a great tool for assessing
freezer performance and troubleshooting any problems. The
memory can store 30,000 events, an estimated 10 years of storage
capacity. The data includes time-stamped temperatures, LN2
level, liquid usage, and any alarms or events.
Password Security
- The multilevel security system comprised of up to ten userspecific
programmable passwords and four security levels can
be customized to grant or restrict personnel access to certain
menus and settings.
Communication Capabilities
- Two independent ports can be used to communicate with other
TEC 3000 controllers, a remote PC, serial printer, or other RS-
485 networks and devices. Options include ASCII, MOD BUS,
Printer and One Fill All Fill (OFAF).
Regular Price $3,173.00 Special Price $2,697.00
-15%
SKU
21837414S
Manufacturer
MVE Biological Solutions
Specifications